The Bebur BT6308-CD Online Chlorine Dioxide Analyzer For Water sensor adopts a membrane that is less affected by chlorine, featuring stable performance, low maintenance and long service life.
The Bebur BT6308-CD Online Chlorine Dioxide Analyzer For Water sensor adopts a membrane that is less affected by chlorine, featuring stable performance, low maintenance and long service life.

▪ Online monitoring of chlorine dioxide concentration applicable to the disinfection and sterilization processes of drinking water, polluted industrial process water and pure water, such as:
▪Precise chlorination control for municipal water quality disinfection and secondary water supply disinfection
▪Chlorination disinfection control of water used in food factories and beverage factories
▪Disinfectant control during the cleaning process of fruits and vegetables
▪Water quality monitoring of medical sewage and other wastewater
▪The sensor adopts the diaphragm amperometric principle
▪Stable performance and accurate measurement
▪No reagents or buffers are required
▪It is not affected by cleaning agents in water
▪It is less affected by residual chlorine
▪Maintain it once every six months
▪It is applicable to most fields of drinking water and process water for process control
The membrane-type chlorine dioxide sensor is a dual-electrode system of membrane-type amperometric polarography, which eliminates zero drift by increasing the current potential. No reagents or buffers are used, and its calibration method is one-point calibration (no need to calibrate the zero point).
The Bebur Online Chlorine Dioxide Analyzer For Water is suitable for occasions where high reliability and ease of operation are required.
Sensor | Membrane amperometric polarographic dual-electrode sensor |
Sensor model | BWCD10.1 |
Application | Water treatment |
Type | Membrane amperometry |
Measurement | Chlorine dioxide |
Range | 0.005-0.5mg/L; 0.005-2mg/L; 0.05-5mg/L; 0.05-10mg/L; 0.05-20mg/L;0.5-200mg/L |
Resolution | 0.001 mg/L,0.01 mg/L,0.1 mg/L |
Repeatability | <1% |
Working temperature | 0-50℃ |
Temperature compensation | Automatic compensation is achieved through an integrated temperature sensor |
Work pressure | 0-0.5bar |
Flow rate | 0.5L/min |
pH value range | pH1-pH12 |
Initial polarization time | About 1 hour |
Response time | The T90 lasts approximately 60 seconds |
Zero-point calibration | No need |
Calibration | Calibrate using the DPD method every one week to three months |
Membrane cap | It is recommended to replace it every 12 to 18 months |
Electrolyte | It is recommended to replace it every 3 to 6 months (depending on the water quality). |
Interference | O3、CL2 |
Installation method | Open circulation pool |