BT6308-Peroxi Online Hydrogen Peroxide Analyzer

BT6308-Peroxi Online Hydrogen Peroxide Analyzer

The Bebur BT6308-Peroxi Online Hydrogen Peroxide Analyzer adopts a membrane sensor. It does not require reagents and buffers. The measurement is stable and the performance is reliable. It can reduce the maintenance amount for customers and lower the usage cost.

Product Details

BT6308-Peroxi Online Hydrogen Peroxide Analyzer

The Bebur BT6308-Peroxi Online Hydrogen Peroxide Analyzer adopts a membrane sensor. It does not require reagents and buffers. The measurement is stable and the performance is reliable. It can reduce the maintenance amount for customers and lower the usage cost.

BT6308-Peroxi Online Hydrogen Peroxide Analyzer

Typical applications

▪The BT6308-Peroxi Online Hydrogen Peroxide Analyzer is applicable to the online monitoring of peracetic acid concentration in the disinfection and sterilization processes of drinking water, sewage, industrial process water and pure water, such as:

▪Dosing control for bottle washing machine

▪CIP Factory application

▪Food cleaning water bleaching system

▪Pharmaceutical factories or the field of pharmaceutical intermediates

Performance characteristics

▪The sensor adopts the diaphragm amperometric principle

▪Stable performance and accurate measurement

▪No chemical reagents are needed

▪It is not affected by the cleaning agents in water

▪No zero-point calibration required

▪It has a relatively long service life

▪Suitable for drinking water, process water and brine

Working principle

The hydrogen peroxide sensor is a diaphragm type dual-electrode system. This electrode can automatically adjust the potential to eliminate zero drift. This design enables water quality testing without the need for additional reagents or buffers, reducing the user's usage cost. During calibration, only high-point calibration is required (zero-point calibration is not necessary). The Online Hydrogen Peroxide Analyzer of Bebur can perform automatic dosing process control. It can control the dosage of H2O2(hydrogen peroxide) by automatically adjusting the flow rate, pump speed or valve position to maintain the concentration of hydrogen peroxide stable at the preset value. Automatic dosing can reduce reagent costs and improve the level of process control.

Technical parameters

Sensor

Diaphragm type amperometric polarographic dual-electrode sensor

Sensor model

BWWP10

Range

0.5-200mg/L, 5-2000mg/L, 50-20000mg/L, 500-50000mg/L, 500-100000mg/L

Resolution

0.1mg/L,1mg/L,10mg/L,100mg/L(It depends on the range of the sensor)

pH range

pH2-pH11

Temperature range

0 to 45℃ (No ice crystals in the measured water)

Temperature compensation

It is automatically completed by the integrated temperature sensor

Pressure range

0-0.5 bar, no pressure shock or vibration

The first use of polarization time

Five hours

Response time

T90 takes about 8 minutes

Flow velocity requirement

Approximately 0.5 liters per minute

Material

PVC-U, stainless steel, silicone

Interference

Chlorine, peracetic acid, ozone, sulfides, phenolic substances >3%

size

Length: 190mm, diameter: 25mm

Diaphragm

It is recommended to replace it every 12 to 24 months (depending on the water quality).

Electrolyte

It is recommended to replace it every 3 to 6 months