Water Hardness Analyzer BC300 A dosing pump meters reagent into the sample, a photometric cell detects the reaction end point, and hardness is calculated from the volume added and reported as CaCO₃. Seven reagents split the range into intervals from 0.21–2.14 to 53.4–534.0 mg/L — pick the one that covers the control band of the water you monitor.

Titrimetric colorimetric 0.21–534.0 mg/L ±5% FS Modbus RS485 7″ touch screen

Hardness is not a parameter where "a rough idea" will do. Softener breakthrough, boiler feedwater excursions and cooling-water scaling usually happen inside the weekly sampling interval of a laboratory. The value of an online hardness analyzer is not a more accurate reading — it is cutting "how often you find out" from a week down to one measuring cycle.

Parameters quoted from the BC300 specification page of Bebur's Water Hardness Analyzer brochure. Supply parameters follow the order-specific technical agreement and factory calibration certificate.

BC300 · softener outlet Measurement 3
0.02mmol/L
= 1.0 mg/L as CaCO₃ · Status: pass
END POINT START T
Interval8 min
RangeT01
Last check−1.2 %

Interface illustration only — not live field data.

Measuring range0.21–534.0 mg/L · 7 ranges
Accuracy / repeatability± 5 % FS
Measuring interval5 min minimum
Comms / outputRS485 · 0/4–20 mA
01 / WHAT IS MEASURED

Hardness is the sum of calcium and magnesium — yet standards write it in different ways

Total hardness is the combined content of calcium and magnesium ions. China's drinking water standard expresses it as calcium carbonate (CaCO₃) in mg/L; the industrial boiler water standard uses mmol/L. The same hardness therefore reads 450 mg/L in one document and 0.030 mmol/L in another — a factor of 50 apart. The first task in online monitoring is not choosing an instrument, but pinning down the unit and the basis for judgement.

Hardness is not directly a toxicity issue, but it decides three things: whether scale forms, whether soap will foam, and when a softener breaks through. In industrial service the most immediate consequence of excess hardness is a calcium carbonate layer on heat-transfer surfaces — every additional 1 mm of scale cuts heat transfer enough to show up plainly in the energy bill; in a boiler it also means local overheating of the heating surface and a higher risk of tube corrosion. That is why boiler water standards are far stricter on make-up hardness than drinking water standards.

Comparison of hardness limits. Only the clauses relevant to selecting this product are excerpted; the complete requirements are in the standards themselves.
ApplicationStandardHardness requirementUnit basis and recommended range
Drinking water GB 5749-2022, Standards for drinking water quality ≤ 450 mg/L As CaCO₃; mandatory national standard. 450 mg/L falls in T07 (53.4–534.0)
Steam boiler · make-up water, external treatment GB/T 1576-2018, Water quality for industrial boilers ≤ 0.030 mmol/L ≈ 1.5 mg/L; falls in T02 (0.005–0.053 mmol/L)
Steam boiler · feedwater, internal treatment GB/T 1576-2018, Water quality for industrial boilers ≤ 4 mmol/L ≈ 200 mg/L; falls in T06 (26.7–267.0)
Hot-water boiler · external treatment GB/T 1576-2018, Water quality for industrial boilers ≤ 0.6 mmol/L ≈ 30 mg/L; falls in T04 (5.34–53.4)
Cooling water / CDU make-up No mandatory national limit By process control Set by the equipment vendor and cycles of concentration; choose the range from the actual band

Conversion basis: 1 mmol/L hardness (as CaCO₃ equivalent) = 50.04 mg/L. Values marked "≈" are conversions for reference only; formal judgement should use the unit specified in the standard.

02 / MEASURING PRINCIPLE

Titrimetric colorimetry, in essence, replaces a pair of hands in the lab with photometric detection

The BC300 follows the titrimetric colorimetric route: a dosing pump meters reagent into the sample on a set programme, the reagent reacts with calcium and magnesium ions and the colour of the solution changes; a photometric sensor tracks absorbance until the reaction end point is reached; the controller then calculates hardness from the volume of reagent already added. It belongs to the same complexometric titration family as the EDTA disodium titration in GB/T 5750.4-2023, with the end point judged photometrically and the dose controlled by a pump.

STEP 01

Sample enters the reaction cell

At the set interval, sample is drawn into the reaction cell and the line is flushed with sample first, so residual volume from the previous cycle cannot dilute this reading.

STEP 02

Dosing pump meters reagent

The dosing pump delivers reagent into the sample under programme control; the dose is governed by pump stroke and frequency, and that volume is the basis for the later calculation.

STEP 03

Complexation and colour development

The reagent complexes with calcium and magnesium while the indicator shows a wine-red colour. Colour intensity corresponds to the total calcium and magnesium still to be consumed.

STEP 04

Photometric end point

The photometric unit monitors absorbance continuously; the point of abrupt colour change is the end point. Detection is not subject to the subjective differences of matching a colour chart by eye.

STEP 05

Calculate and display

The controller converts the accumulated reagent volume into a hardness value, shows it on the 7-inch touch screen, logs it and sends it to the host system.

Why the colour changes at the end point

The complex formed between the indicator and calcium/magnesium is less stable than the complex formed with the reagent. As reagent is added, it first takes the free calcium and magnesium in solution; past the equivalence point it also takes the calcium and magnesium bound to the indicator, which then returns to its own colour — that is the "wine red to pure blue" moment of a visual titration.

Ca/Mg + indicator Start · wine red Reagent complexes Transition Indicator free End · pure blue

What each of the three common methods can and cannot do

Instruments marketed as "online hardness analyzers" do not share one principle. Knowing which class a unit belongs to matters more than comparing accuracy figures.

MethodPrinciple and basisWhere it fitsLimitations
Titrimetric colorimetric
used by BC300
Metering pump doses reagent, photometric detection of the end point; same complexometric titration family as the EDTA disodium method in GB/T 5750.4-2023 Drinking water, boiler water, cooling water, food and beverage and medical water — anywhere the result must be comparable with the laboratory Requires reagent; discrete measurement (5 min minimum); reagent must be replaced on schedule
Photometric colorimetry Indicator develops colour, absorbance is read and converted via a calibration curve Raw and process water with high or widely varying hardness, trend monitoring High concentrations need dilution; colour and turbidity directly bias the reading
Ion-selective electrode (ISE) Potential response of a calcium or combined calcium/magnesium electrode Continuous trend monitoring where reagent-free maintenance is required Weak stability at low concentrations, frequent calibration; interference from coexisting ions

Selection tip: if hardness results must line up with third-party laboratory reports, choose a titrimetric method; if only the trend of "over or under limit" matters, colorimetric and ISE units are adequate. Write the method into the procurement technical agreement, so a mismatch in basis does not surface only at acceptance.

03 / SPECIFICATIONS

BC300 technical data

The table below is taken from the BC300 specification table in Bebur's Hardness Analyzer brochure. The range is not a single figure: the instrument covers 0.21–534.0 mg/L overall, while the usable interval is set by the reagent range selected — see the next table.

ParameterBC300Selection note
Measuring principleTitrimetric colorimetricDosing pump adds reagent, photometric end point
Measuring range0.21–534.0 mg/L (ppm) CaCO₃Segmented across seven reagent ranges
Accuracy±5 % FSFS = full scale, not reading
Repeatability±5 % FSSame basis, referred to the full scale of the range in use
Measuring interval5 min minimum, several optionsShorter intervals consume reagent faster
Display7-inch LCD touch screenHistory curves and alarms on screen
Dimensions300 × 300 × 180 mmLeave room for reagent replacement
Weight4 kgCheck wall or cabinet load capacity
Temperature range5–45 °CHot samples need temperature conditioning
Humidity range20 %–95 %, indoorOutdoor siting needs a protective cabinet
Pressure range0.2–2 barAbove 2 bar, reduce pressure before the cell
Protection classIP54Dust and splash protected, not for wash-down areas
CommunicationModbus RS485Optional WiFi / 4G wireless
Analog output0/4–20 mAConfirm the scaling expected by the host device
Digital input1 channelFor external triggering or status capture
Digital output2 channelsAvailable as alarm or interlock contacts
Data storageMore than 1,000,000 records (32 G)At 5-minute intervals, coverage far exceeds 10 years

Note: accuracy and repeatability are referred to full scale (FS). On the T01 range (0.21–2.14 mg/L), ±5 % FS is about ±0.11 mg/L; on the T07 range (53.4–534.0 mg/L) it is about ±27 mg/L. When judging a tight limit, select the range that sits closest to that limit.

Seven reagents that split the range

One BC300 covers different concentration intervals by changing reagent range. The brochure already gives all four unit columns side by side, so values can be checked directly against the standard text without conversion on site.

Reagent mg/L °dH °f mmol/L
T010.21–2.140.012–0.120.021–0.2140.002–0.021
T020.53–5.340.03–0.30.053–0.5340.005–0.053
T031.60–16.00.09–0.90.160–1.6020.016–0.160
T045.34–53.40.3–3.00.534–5.3400.053–0.534
T0516.02–160.20.9–9.01.602–16.020.160–1.602
T0626.7–267.01.5–152.670–26.700.267–2.670
T0753.4–534.03.0–305.340–53.400.534–5.340

Reagent codes are WAR-Hardness-300T01 through T07 and are used with the BC300 controller. The mg/L column is expressed as CaCO₃; values are taken from the reagent information table in the brochure. Once a range is selected, the alarm threshold must sit inside that range, otherwise an over-limit condition can occur without being measurable.

Range selection example: the limit for steam boiler make-up water with external treatment is 0.030 mmol/L (about 1.5 mg/L). The upper limit of T01, 0.021 mmol/L, does not cover it, so T02 (0.005–0.053 mmol/L) is the right range. Selecting T03 or higher puts the limit below the lower end of the range, and the reading loses any discriminating value.

04 / UNIT CONVERSION

One hardness value, six ways to write it

Hardness is expressed differently across standards and industries. Enter any one value and the others update instantly, along with a note on which BC300 reagent range that value falls into. The conversion factors are fixed and can be used directly to check figures in technical agreements and acceptance sheets.

mg/L (as CaCO₃)
450.0mg/L
mmol/L
8.99mmol/L
°dH, German degree
25.2°dH
°f, French degree
45.0°f
°e, English degree
31.5°e
ppm
450.0ppm
Conversion basis: 1 mmol/L = 50.04 mg/L; 1 °dH = 17.848 mg/L; 1 °f = 10 mg/L; 1 °e = 14.3 mg/L. °dH, °f and °e are customary units — convert back to the unit specified by the standard for formal judgement.

Reference: the total hardness limit in GB 5749-2022 is 450 mg/L (as CaCO₃), equal to 8.99 mmol/L or 25.2 °dH, which falls inside the BC300 T07 range (53.4–534.0 mg/L).

05 / APPLICATIONS

Six places that genuinely need continuous hardness readings

Whether continuous monitoring pays off depends on the cost of an excursion and the delay before anyone notices it. What these applications share: once hardness breaks through, the consequences start accumulating within hours, while manual sampling intervals are typically a day to a week.

01

Drinking water and municipal supply

Hardness at the plant outlet and at the network extremities affects taste, scaling in pipes and the load on downstream softening. The T07 range covers the 450 mg/L limit of GB 5749-2022, so readings can be checked directly against published water quality reports.

02

Boiler water and softener outlet

The limit for make-up water with external treatment is 0.030 mmol/L (about 1.5 mg/L), which calls for the T02 range. Put the sampling point after the softener and before the deaerator, on a stable flow section, so breakthrough shows up within the next measuring interval.

03

Cooling towers and cycles of concentration

Hardness rises with cycles of concentration and is a direct basis for judging scaling risk and blowdown timing. Used together with conductivity and pH, it describes the stability index of the water far more accurately.

04

CDU make-up and secondary loop water in liquid cooling

Cold-plate channels are narrow, so a calcium carbonate layer is amplified in both heat transfer and flow resistance. Make-up hardness is a must-test item when qualifying a make-up water source, and belongs to the cooling water monitoring family.

05

Food and beverage process water

Hardness affects formulation stability, taste and cleaning performance. Online readings serve as continuous evidence for compliance and process adjustment, not just a post-shift grab sample record.

06

Medical facility water

Cleaning and disinfection, laboratory work and haemodialysis all set explicit hardness requirements. Online monitoring documents continuity of water quality and reduces reliance on a single sample.

06 / INTERFERENCE AND INSTALLATION

Most field deviation comes from sampling, not from the instrument

The same instrument can differ by a factor of two between two sampling arrangements of the same water. The items below are the ones most often overlooked during commissioning, and the easiest to correct.

InterferenceEffect on the readingEngineering remedy
Turbidity and colour Scattered light shifts the photometric baseline, so the end point comes early or late Add filtration ahead of the cell; sample downstream of the filter
Iron, manganese, copper, zinc Compete with the indicator for complexation, smearing the end point or inflating the reading Add a masking agent; for water high in iron, sample after iron removal
pH excursion Strong acid or alkali weakens buffering and shifts the indicator's colour-change range Match the reagent dose setting to the alkalinity of the sample
Sample temperature Affects reaction rate and the sharpness of the colour change Keep within 5–45 °C; cool hot samples first
Air bubbles and flow loss An inaccurate sample volume produces a direct proportional error Use overflow sampling; avoid high points that trap air and dead legs in the line
Phosphate, silicate In chemically dosed boiler samples these may precipitate with calcium and consume titrant Adjust the sampling point or method around the dosing regime; cross-check with another method if needed

Pre-commissioning checklist

Sampling point

Avoid the immediate downstream of a dosing point, gas–liquid mixing zones and branches that rarely flow; keep the sample line short with few bends.

Range and threshold

Choose the reagent range from the control band of the water being monitored (T02 for boiler make-up, for example), then set the alarm threshold inside that range and below the limit.

Cross-checking

In the first month after commissioning, compare with a laboratory titration weekly. If deviation exceeds the repeatability figure, check sampling and reagent before touching the calibration factor.

Reagent and wear parts

Log reagent batch numbers and opening dates; inspect pump tubing and lines quarterly so expired reagent does not introduce a systematic bias.

What this page does not cover: this instrument measures total hardness in conventional industrial and municipal water. Trace hardness in high-purity water (ultrapure or semiconductor-grade), samples containing large amounts of organic solvents or strong oxidants, and applications that must separate calcium hardness from magnesium hardness all call for a dedicated method or instrument, and readings from this unit should not be used as the basis for judgement.

07 / FAQ

Eight questions from selection and operation

Are results from the BC300 titrimetric colorimetric method comparable with laboratory EDTA titration?
The BC300 uses titrimetric colorimetry: a dosing pump adds reagent that reacts with calcium and magnesium ions, a photometric cell detects the reaction end point, and hardness is calculated from the volume of reagent added. It belongs to the same complexometric titration family as the EDTA disodium method in GB/T 5750.4-2023; the difference is that the end point is detected photometrically and the dose is set by a pump rather than by eye and a burette. Comparability depends on controlling the sampling point, sample temperature and reagent batch — compare against a laboratory titration weekly during the first month of operation.
The range reads 0.21–534.0 mg/L — is that one range or several?
It is covered in segments. Seven reagents (WAR-Hardness-300T01 through T07) split 0.21–534.0 mg/L into seven intervals: the lowest, T01, is 0.21–2.14 mg/L (0.002–0.021 mmol/L), and the highest, T07, is 53.4–534.0 mg/L (0.534–5.340 mmol/L). Choose the reagent by where the control band of your water falls, not by the instrument's overall upper limit.
Accuracy and repeatability are quoted as ±5 % FS — how much error is that in practice?
FS means full scale, not reading. On the T01 range (0.21–2.14 mg/L), ±5 % FS is about ±0.11 mg/L; on the T07 range (53.4–534.0 mg/L) it is about ±27 mg/L. So when judging a tight limit, choose the reagent range that sits as close as possible to that limit rather than measuring a low concentration on a wide-range reagent.
The boiler make-up water limit is 0.030 mmol/L — can this instrument measure that?
0.030 mmol/L is about 1.5 mg/L as CaCO₃, which falls inside the T02 range (0.005–0.053 mmol/L, roughly 0.53–5.34 mg/L). The upper limit of T01, 0.021 mmol/L, does not cover it. So for steam boiler make-up water, configure T02 and set the alarm threshold below the limit — 0.025 mmol/L, for example — so an alarm appears early in the breakthrough.
With a 5-minute cycle at best, can it drive softener regeneration interlock?
The measuring interval can be set to a minimum of 5 minutes, and the two digital outputs can serve as alarm or interlock contacts. For softener outlet monitoring, a 5-minute response covers most breakthrough events; but if the regeneration programme itself must act within seconds, pass the instrument alarm to the control cabinet as one trigger condition and let the cabinet run regeneration with flow or time logic rather than driving it directly from the analyzer.
Which samples interfere with the hardness measurement?
High turbidity, pronounced colour, heavy metals such as iron, manganese, copper and zinc, strong acid or alkali, and samples with a lot of phosphate or silicate all interfere with colour development and the photometric end point. In practice these are handled with pre-filtration, masking agents and temperature control; where interference cannot be removed, switch to another method or sample after pretreatment. The BC300 is specified for 5–45 °C, so out-of-range samples need temperature conditioning first.
How much data can it store, and can it connect to our existing PLC?
Internal 32 G storage holds more than one million test records; at 288 measurements per day (one every 5 minutes) that covers far more than 10 years in theory, with actual capacity depending on what is logged. Externally it provides 0/4–20 mA analog output and Modbus RS485, optional WiFi / 4G wireless, plus one digital input and two digital outputs, so it can connect directly to a PLC, DCS or softener control cabinet.
What are the requirements for the installation environment?
The unit measures 300×300×180 mm and weighs 4 kg, is rated IP54 and is designed for indoor use at 20 %–95 % humidity, with sample pressure 0.2–2 bar and ambient temperature 5–45 °C. Keep it out of direct sunlight, strong vibration and corrosive gases; keep the sample line short, avoid high points that trap air, and leave room for reagent replacement.
08 / STANDARDS AND SOURCES

Where the figures on this page come from

The limits and method names on this page are quoted from the public texts of the standards listed below. Standards may be revised — check the version currently in force before relying on a limit.

01
GB 5749-2022, Standards for drinking water quality — total hardness (as CaCO₃) limit 450 mg/L. Issued 2022-03-15, effective 2023-04-01.
National standard full-text system · official PDF
02
GB/T 1576-2018, Water quality for industrial boilers — hardness limits (mmol/L) for feedwater, make-up water and boiler water of steam and hot-water boilers.
Standard text · special equipment information platform
03
GB/T 5750.4-2023, Standard examination methods for drinking water — Part 4: sensory and physical indices; total hardness is determined by EDTA disodium titration. Issued 2023-03-17, effective 2023-10-01.
04
GB/T 7477, Water quality — Determination of the sum of calcium and magnesium — EDTA titration method; shares the same titration system as GB/T 5750.4 and is commonly used for hardness in non-potable matrices.
05
ISO 6059, Water quality — Determination of the sum of calcium and magnesium — EDTA titrimetric method; the international standard for the same method.
06
GB/T 50050-2017, Code for design of industrial recirculating cooling water treatment; basis for water quality control items and indices in recirculating cooling systems — specific limits depend on process and cycles of concentration.
07
Bebur, Water Hardness Analyzer product brochure, BC300 specification table and WAR-Hardness-300T01–T07 reagent information table. All instrument specifications and range segmentation on this page are quoted from that page.
09 / ABOUT THIS PAGE
Written by
Bebur application engineering team
Water analysis product line
Technical review
Technical lead, water analysis product line
First published / updated
2026-11-06
Content policy
Instrument specifications quoted from the product brochure;
limits quoted from the standards; derivations and conversions flagged separately

This page is a selection reference. It does not replace a site survey, process calculation or contractual technical agreement. Instrument specifications and range segmentation are quoted from the BC300 product brochure; supply parameters follow the order-specific technical agreement and factory calibration certificate. The readout shown on this page is an illustration, not live field data.

Confirm the range, interval and alarm threshold for your water

Send the sample conditions (temperature, pressure, expected hardness band, interfering ions) and the sampling point, and we will return the matching reagent range and the technical datasheet.